World First Hybrid Pyrolyzer

Maximize the benefit of Hybrid
Two heating methods by “Curie Point” and “Furnace” are both available in one system. It is the most versatile pyrolyzer that offers 3 different analysis modes.

JHI-08 Controller and Oven unit

Pyrolysis Mode
Two ways of pyrolysis are available. You can choose Curie Point pyrolysis using Pyrofoils or Furnace pyrolysis with ceramic heater depending on sample shapes and target.

Hybrid Mode
You can conduct thermal extraction and Curie Point pyrolysis on one sample continuously without interruption.

EGA Mode (Evolved Gas Analysis)
You can program multiple phases of temperature elevation with different duration and rate. It is usually directly coupled with a mass spectrometer to find thermal decomposition behavior of samples.

Easily removable module prevents cross contamination in the flow chancel
During pyrolysis, approximately 60% of tar-like high-boiling materials remain in the sample tube. This is especially a problem for samples of molded resins and rubbers that contain inorganic filler, carbon powder and other additives. JHI-08 has easily replaceable sample tube and needle assy for cleaning . This minimizes cross contamination and results in better and reproducible chromatograms.

Options to make it more versatile

Auto sampler unit APC-50 (Click here for details)
APC-50 Auto sampler automates the sample feeding/collection process for both Pyrofoils and sample cups. Magnetic sample retriever does not consume Helium gas. It is quite easy to attach to and remove from pyrolyzer. APS-50 Auto sampler for Pyrofoils only is also available.

Pyrolysis in air atmosphere and Column inlet cooler
Purge Gas Selector (PGS) allows to switch carrier gas to air easily. By combining with Column Inlet Cooler (CIC), you can perform pyrolysis under ambient air.

Reanalysis adsorption set

The generated gas is split again (2nd splitter) after GC injection port. One goes to GCMS for standard analysis, the other is vent out and collected to mini-PAT (Sample tube with TenaxGR and 280°C Pyrofoil). The trapped gas in mini-PAT can be reanalyzed using JCI-77.

APC-50

Specifications

 

 Curie point pyrolysis (CP-PY)
 Heating method Curie point heating
 Sample setting by dropping a Pyrofoil in the oven
 Pyrolysis temp. by choosing a Pyrofoil
Temp. range from 160°C to 1040°C
 Pyrolysis time 1 to 99 sec
 Needle 0.64 mm OD x 70 mm
 Sample tube material Quartz glass
 Dimentions, Weight W120 x D121 x H328 mm, 2 kg
 Furnace pyrolysis (F-PY)
 Heating method Furnace oven heating
 Heater Tubular ceramic heater
 Sample setting by dropping Pyrofoil or sample cup in the oven
or lower the holder by hand
 Pyrolysis temp. 40 to 800°C by 1°C unit
 Temp. elevation rate Max 600°C / min
 Needle 0.64 mm OD x 70 mm
 Sample tube material Quartz glass
 Dimensions, Weight W120 x D121 x H328 mm, 2 Kg
Controller unit
 Needle heater 40°C – 300°C
 Oven heater Ceramic heater 40°C – 800°C
 Radio frequency generator Radio frequency output : 48W
Fall detect sensor is installed
 Carrier gas GC/JHI Switching valve is installed
 Dimensions, Weight W180 x D500 x H390 mm, 16 Kg
 Power 100 – 220V, 500VA (Max)

 

Auto Sampler for Hybrid Pyrolyzer JHI-08

 

Auto Sampler APC-50
APC-50 Auto sampler automates the sample feeding/collection process for both Pyrofoils and sample cups. Magnetic sample retriever does not consume Helium gas. It is quite easy to attach to and remove from pyrolyzer. APS-50 Auto sampler for Pyrofoils only is also available.

Two ways of sample feeding
Having two types of turntable, APC-50 supports sample feeding for both Pyrofoils and sample cups. Two turntables come as a standard accessory.

Turntable for Pyrofoils

Turntable for sample cups

Pyrofoil and Cup sampler APC-50
Maximun Number of samples 50 samples at a time
Sample introduction By turntable
Sample Retrieval By magnetic retriever stick
Air Purge Auto purge by solenoid valve
Dimensions, Weight W175 x D265 x H662 mm
including JHI-08 Pyrolyzer, 8 kg

 

APC-50 installed on JHI-08

Pyrofoil Sampler

 

Auto Sampler APS-50
APS-50 Auto sampler automates the sample feeding/collection process for Pyrofoils (Sample cups are not supported). Magnetic sample retriever does not consume Helium gas. It is quite easy to attach to and remove from pyrolyzer. APC-50 Auto sampler for Pyrofoils and sample cups is also available.

Sample feeding
Maximum 50 pyrofoils can set at a time.

Foil magazine

Pyrofoil and Cup sampler APS-50
Maximun Number of samples 50 samples at a time
Sample introduction By turntable
Sample Retrieval By magnetic retriever stick
Air Purge Auto purge by solenoid valve
Dimensions, Weight W275 x D272 x H612 mm
including JHI-08 Pyrolyzer, 6 kg

 

APS-50 installed on JHI-08

Pyrolyzer

 

 Hybrid Pyrolyzer
Curie Point Pyrolyzer