Production Scale Model

Production Scale Model
Selectable 3 models with maximum flow rate at 1,400 mL/min, 2,300 mL/min and 4,000 mL/min.

High performance pump
Reciprocating triple plunger delivers stable flow rate and recycling performance.

Recycling preparative chromatograph (patented)
The system is equipped with non-diffusion recycling system (patented) that prevents the sample from running back into solvent line, baseline shifting, broadening and damaging detectors with back pressure. It protects detectors, columns and your precious samples.

Gradient system
Binary high pressure gradient mode is supported. In addition to mobile phase supply, a separate solvent line to wash column is standard-equipped.

Repeat injector
Enables to auto-inject large volume of sample successively.

PC control
Enables to program operations from sample injection to fraction collection and run it remotely and automatically.

 


Solvent saving system
Utilizes the solvent consumed waiting for peak elution and saves large amount of solvent into dedicated container.

Specifications

 

Model LC-Prep1000 LC-Prep2500 LC-Prep4000
 Pump Model P-1000 P-2500 P-4000
 Pump Type Reclprocating triple plunger
 Gradient High pressuer gradient with 2 solvents
 Flow Mode Constant flow
 Pressure Display & Limiter 15 MPa 9.4 MPa 11.7 MPa
 Flow Rate  1400 mL/min  2300 mL/min  4000 mL/min
 Injection Method Repeat Injector
 Sample Injection Amount 25 – 1,000 mL 60 – 2,500 mL 80 – 5,000 mL
 Collection Mode Peak and Time program mode
 Auto run Save previous operation and rerun up to 99 times
 Nozzle of Frction 5
 Controller Unit Type PLC touch panel, PC control
 Sample Injection Interval 99 hours (max)
 Sample Injection No. 99 times (max)
 Recycle Time Control Start and End time :1 to 999 min.
 Recycling NO. 99 times (max)
 Recycle System With non-Diffusion function Recycling system (patented)
 Safty Divice Prevent from Electric and Liquid leakage, pressure, high temperature
 Dimension(mm) 1050(W) x 1400(H) x 880(D) 1050(W) x 1400(H) x 1000(D)
 Power Dimension 580(W) x 1150(H) x 560(D) mm
 Power AC200 – 240V, 3000VA
Weight Main body approx. 400Kg, Power approx. 100Kg

Dynamic Axial Compression Column

Please ask for details.

Model Dimensions Max Bed Length
DAC100-700 100 mm ID x 700 mm 400 mm
DAC150-700 150 mm ID × 700 mm 400 mm
DAC200-700 200 mm ID × 700 mm 400 mm
DAC250-700 250 mm ID × 700 mm 400 mm
DAC300-700 300 mm ID × 700 mm 400 mm