LaboACE LC-5060 Plus II

Hızlı Bağlantı

Ürünlerimiz Hakkında Bilgi Alın

Ürünlerimiz hakkında her türlü sorunuz için temsilcilerimizle iletişime geçebilirsiniz.

Ürün Katalogu

Katalog.pdf

Talep Dosyası.docx

Servis Talebi Oluştur

Ürünleriniz için kurulum, bakım ve arıza gibi konularda servis talebi oluşturabilirsiniz.

Servis Talep Formu

Standard Model

Manual Collection
LaboACE LC-5060 Plus II is a most popular Recycling Preprative HPLC system that dedicated to manual fraction collection. Start and stop of peak recycling separation can be made easily as desired just by a single click on “Recycle” button on the touch panel. LaboACE LC-5060 Plus II promises to make your purification steps simple, quick and easy, especially when separating different samples every time.

Small to fit inside a fume hood or on a benchtop
From injector to fraction collector, all necessary components are placed in the compact body. It is so compact as A3 size paper (two letter-size papers) that it is easy to set up inside a fume hood or place on a benchtop.

Recycling preparative chromatograph (patented)
The system is equipped with non-diffusion recycling system (patented) that prevents the sample from running back into solvent line, baseline shifting, broadening and damaging detectors with back pressure. It protects detectors, columns and your precious samples.

Auto Cleanup Function
It cleans inside the flow channels automatically. Auto cleanup function along with the patented non-diffusion recycling system prevent system residual contamination so that you can obtain ultra pure compounds with confidence. Solvent replacement can also be made much easily.

Reliable Solvent Delivery
With more than 50 years’ experience of developing recycling technology, the system has JAI-made serial double plunger pump optimized for stable delivery and recycling performance. The recycling efficiency is greatly improved.

Right Fraction Timing
The time difference due to delay volume from sample detection to fraction is automatically calculated.

  

Compact body with minimized dead volume
Minimized internal dead volume and the newly developed non-diffusion recycling system prevent base line shifting and peak broadening.

Simple and Intuitive Operation interface

Touch screen allows anyone to operate LaboACE easily on the screen from solvent delivery to fraction.

Improved the ease of operation
Various functions are installed to use the instrument conveniently, such as pump stop timer that stops pump at desired time and bottle capacity that prevents fractionation from overflowing from vials.

Detectors

Two detectors can be fitted in the LaboACE series small body and can detect simultaneously.

UV Detector

Especially designed for recycling preparative HPLC with stability and high detection sensitivity.

UV-254 LA UV-VIS 4ch 400LA UV-VIS 4ch 800LA
 Wavelength Fixed at 254 nm
Select wavelength as option
4ch variable
from 200 – 400 nm
4ch variable
from 200 – 800 nm
 Measurement method Single beam
 Flow cell Optical path length: 0.3 mm Volume: 45 μL
(Option : 1.4 mm and 2.4 mm)

Refractive Index Detector

It gives quick baseline stabilization and won’t require much time before injection.
Applicable for both preparative and analytical HPLC.

 RI-700 LA
 Measurement method Deflection
 Flow cell Volume 8 μL
 RI range 1.00 – 1.75

Specifications

 

  LaboACE LC-5060 Plus II
 Pump P-60LA2
 Solvent delivery system Reciprocating double plunger
 Flow rate range 0.1 – 60.0 mL/min
 Maximum discharge pressure 20 MPa
 Plunger  7.0 mm ID
 Stroke  8 mm
 Discharge rate per stroke  300 μL
 Gradient module Optional : Low pressure binary gradient.
It supports mixing two solvents isocratically for recycling separation.
 Variable injector (Manual)  Rheodyne 7725i
 Sample loop  10 mL as standard
 Repeat injector Optional
 Loop injector Not available
 Recycle system  With non-diffusion function recycling system (Patented)
 Recycle method Switching scheme of solenoid valve
 Auto program
・Event programming
・Auto fraction collection
・Auto replay
・Stacked injection
Not available
 Fraction collector Built-in single fraction nozzle
 Touch screen 11.6 inch touch screen*
 Pressure display and limiter Display and program on LCD touch screen
 Event signal  Injection, Recycle and Collect
 Auto cleanup To wash flow channel by auto program
 Pump stop timer  To stop pump at 0-999 mins
 Bottle capacity To stop pump at 0-999 mL when collecting and prevent from overflow
 Timer with alarm To set alarm at 0-999 mins
 Dimensions 280(W) x 445(H) x 420(D)
 Power supply   100V, 500VA/200-240V, 500VA
 Weight 18 Kg
 Column stand Option
 Solvent stand Option
  Control Unit
 Touch Screen Interface 11.6 inch touch screen
 Operating system Windows 10
 Preinstalled software JAI Scan for chromatographic data acquisition (4ch UV and RI)
JAI Control for controlling LaboACE LC-5060 Plus II

 